CFD-driven
Optimization of Room Airflow Patterns and Their Effect on Hood Containment
for Flexible Labs of the Future
Alexy Kolesnikov
and Ray Ryan, Flow
Sciences, Inc.
Douglas B. Walters,
KCP, Inc.
Objectives:
The function of the standard lab has changed. The
past that was rich in synthesis-based R&D has moved into the
new millennium and been replaced with advanced analysis/discovery
processes utilizing highly sophisticated computer technology and
high throughput robotic processes requiring safe, adaptable and
energy efficient containment devices functioning in a fast changing
open lab environment. Two interlinked aspects of containment performance
must be considered to adequately address these demands. First, the
conventional fume hood, which controls about 75% of the lab containment
technology to date, will have to undergo monumental changes in design
aimed at improving containment and energy efficiency at lower face
velocities (low airflow design). Second, the airflow patterns in
the lab itself create cross-drafts near the sash opening and their
effect on hood containment performance must be investigated for
specific lab configurations in order to provide a comprehensive
containment analysis of the entire lab/hood airflow system.
Findings:
Following the results presented at the October 2002 Labs 21st century
conference, that documented performance of a newly designed containment
system combining a rapid recovery variable air volume control mechanism
with computer-optimized laminar airflow pattern inside the work
area, current paper extends the application of this technology to
a new low-airflow hood design. The new hood design is shown to provide
stable containment in the face velocity range of 40-60 ft/min, resulting
in significant operational energy cost reduction. Further, the CFD-driven
airflow modeling is used to investigate hood containment performance
within a sample laboratory room configuration. In that, the effects
of diffuser/hood position, diffuser/hood separation, hood separation,
hood position and equipment arrangement within the lab on the airflow
distribution near the sash opening are parametrically analyzed for
a variety of geometric configurations. Airflow patterns within the
laboratory are shown to directly influence hood containment performance,
with CFD-driven parametric studies enabling to make specific lab
design and hood placement recommendations prior to any construction
installation or mock up.
Labs21 Connection:
Two interlinked aspects affect performance of containment devices
within the lab environment, namely airflow distribution inside the
device itself and airflow patterns inside the laboratory. In that,
inadequate hood design causes turbulent, irregular airflow inside
the hood work area, while improper ventilation design and/or hood
placement within the lab causes significant cross-drafts near the
sash opening of the hood, both leading to loss of containment and
breach of occupant safety. The problem is compounded by the modern
open laboratory flexibility demands. Experimental tests are costly,
provide limited data and often identify the problem only after the
laboratory has been designed and build and hoods installed. Computational
fluid dynamics, i.e. computer modeling of airflows inside the hood
and the lab room environment, permits to parametrically study airflow
distribution for a variety of geometric configurations, optimizing
diffuser, equipment and hood layouts for a particular lab configuration,
thus allowing for predicting and optimizing containment performance
of the entire laboratory/hood airflow system, making changes to
the entire system layout knowing the exact effect of such variations,
using only enough air changes per hour to satisfy the requirements,
conserving energy via hood face velocity optimization, all prior
to construction or mock up. The paper presents results and recommend-ations
for a sample lab room configuration, documents and validates performance
of a newly designed low airflow VAV hood system, thus addressing
the following Labs 21 approaches: use lifecycle cost decision making,
by emphasizing safety and long-term energy savings; employ a broad
range of sustainable energy strategies, by optimizing supply/exhaust
and hood placement strategies and performance and documenting design
of a low airflow VAV hood; promote energy efficiency operation and
training efforts, by introducing an low airflow VAV containment
system as well as computer modeling airflow visualization data absolutely
unavailable otherwise.
Biographies:
Dr. Alexy Kolesnikov
joined Flow Sciences in May 2000 as a Sr. Design engineer. Dr. Kolesnikov
received his PhD in Mechanical Engineering with a focus in fluid
dynamics and heat transfer from the University of Tennessee, Knoxville
in May 2000 and his BS and MS in mechanics and mathematics from
Moscow State University in May 1991. He specializes in advanced
computer simulation of complex multi-dimensional fluid/gas flows
in industrial processes. His research has been published in peer
reviewed scientific journals and presented at numerous CFD-related
conferences.
Raymond F. Ryan, is
the Founder, President and CEO of Flow Sciences, Inc. Flow Sciences
engineers, designs, manufactures and markets solutions for the modern
technology company through the engineering of ventilation containment
systems that protect workers and the environment from exposure to
toxic and "potent" compounds. The products are used to
protect workers from exposure during the handling of hazardous powders
and vapors. These operations take place in laboratories, pilot plants
and manufacturing locations. In 1999 Flow Sciences received an award
for being one of the fifty fastest growing high technology companies
in North Carolina. Prior to founding Flow Sciences, Mr. Ryan founded
two other companies that serviced the pharmaceutical safety industry.
Laboratory Safety Services, which Mr. Ryan sold in 1986, provided
certification testing for biological safety hoods, chemical fume
hoods and laboratory airflow control systems. The customer base
was the major medical schools, pharmaceutical and chemical companies
in metro NY, NJ and Philadelphia. Mr. Ryan also founded and sold
a manufacturers rep company that specialized in capital equipment
for the biomedical research and biotechnology markets.
In the late 60's and early 70's Mr. Ryan worked for Mettler Instruments,
manufacturer of laboratory automation equipment and balances, Arthur
H. Thomas Company, a manufacturer of laboratory equipment and was
a research engineer for Fairchild Environmental Systems developing
solutions for water and air pollution monitoring equipment. From
1966 through 1968 Mr. Ryan served in the US Army as the Director
of Communication Center in Chu Lai, Vietnam for the 3rd BN/82nd
Artillery. He received the US Army Commendation Medal. He received
his BS in Chemistry at Adelphi University, NY and studied Chemical
Engineering at Clarkson University, New York. Mr. Ryan is a member
of: American Chemical Society (ACS), American Society of Safety
Engineers (ASSE), American Biological Safety Association (ABSA),
American Industrial Hygiene Association (AIHA), American Society
of Heating, Refrigeration and Air Conditioning Engineers (ASHRAE),
International Society of Pharmaceutical Engineering (ISPE), Scientific
Equipment & Furniture Association (SEFA). He has authored numerous
journal articles, lectured internationally, and has several patents
pending.
Douglas B. Walters,
Ph.D., CSP, CCHO is on the advisory board for Flow Sciences, Inc.
Dr. Walters is President of KCP Inc.; a consulting company in Raleigh,
NC specializing is laboratory health and safety, laboratory ventilation
and hoods, and industrial hygiene. He is the former Head of Laboratory
Health and Safety for the National Toxicology Program (NTP at the
National Institute of Environmental Health Sciences (NIEHS, NIH)
and former adjunct associate professor in the Program of Public
Health at Old Dominion University. Dr. Walters has been a reviewer
and served on the Editorial Board for several publishers. Dr. Walters
is a Certified Safety Professional (CSP) and a Certified Chemical
Hygiene Officer (CCHO). He is the American Industrial Hygiene Association
(AIHA) Liaison to the American Chemical Society (ACS), a member
of ANSI Z 9.5 Laboratory Subcommittee, a member of numerous AIHA
and ACS committees, and is past Chair of the ACS Division of Chemical
Health and Safety and the ACS Northeast Georgia Section. Dr. Walters
has lectured internationally, received two ACS awards for his contributions
to chemical health and safety, and received three government awards.
He has authored more than 100 publications, books, book chapters,
electronic databases, journal articles, one patent and one video.
Dr. Walters received his PhD in Chemistry at the University of
Georgia, Athens, received his MS and BS in Chemistry from Long Island
University, Brooklyn, NY.
|